Insight
Scientific Imaging: Visual Data and Ethics
By Jerry Sedgewick
Excerpted from Scientific Imaging with Photoshop: Methods, Measurement, and Output (New Riders)
Dateline: September 24, 2008
More Insight articles
In the world of scientific research, images fall broadly
into two categories: the original image and the corrected image.
The original is acquired via an imaging device without any corrections
applied in software. The corrected image, often referred to as
“enhanced,” is often used for presentation and publication.
The distinction is crucial when suspicions arise about content in images.
Because images are used as visual proof of experimental evidence,
certain alterations to the content may be viewed as unethical. Certainly,
any additions to the content from other images, or intentional alterations
of visual data to accommodate the experimental hypothesis, will
always be unethical. The sole means for determining the extent or the
existence of alterations or additions lies in looking at the original.
A kind of reasoning can then follow: If the original is the indisputable
visual evidence for experimental conclusions, only the original should be
used. No alterations to visual data should be applied, and no corrections
should be made for imaging device inconsistencies or shortcomings.
This view operates along incorrect assumptions. Out of many assumptions,
the following two are most often voiced (informally):
- Imaging systems and associated instrumentation can always produce
a visual representation identical to what was seen by eye
(within limitations inherent to two-dimensional renditions of three-dimensional
specimens or scenes) if only those using the instrumentation
were more knowledgeable.
- Problems with images are often a result of poorly prepared
samples or the use of suboptimal dyes and stains: Bench practices
need to be improved.
While both of these assumptions are true in many instances, they
are based on a false premise that the instrumentation by itself, even
when used correctly, produces accurate representations.
Accurate Representation of Visual Data
Efforts must be taken to present visual data as closely as possible to
what was once perceived by eye. In other words, an image offered as
proof must be a true representation of what was once seen. Any deviation
from a correct representation is a misrepresentation, and the
images become varying degrees of inaccurate data.
Accurate representation more often requires post-processing than
not, except in instances in which optical densities or intensities
(OD/I) are measured. Post-processing is required, for the most part,
because of limitations in imaging devices and associated instrumentation.
These limitations include, among others, the use of anti-aliasing
filters in front of light detectors in many cameras, leading to blurring
of images; inclusion of noise in images as a result of approaching
detection limits for instrumentation; and variability in the energy
source. Thus, for any of these reasons, the original image is corrected
to create a better representation of visual data.
Additional reasons for inaccurate representation of the original as a
result of imaging and display devices follow:
- Screen calibration. The screen on which the image is viewed may
not display intensities and gradients of colors and grayscale levels
appropriately (below, left). The gradients/intensities displayed
on the screen may be inaccurate. When these images are incorrectly
displayed and then presented or published, the inaccuracy
is either perpetuated or the colors and contrasts change to even
greater misrepresentations. Post-processing in color- and contrast-managed
programs (Photoshop) will lead to better representations
than the original (below, right).

- Hue shift. If a digital camera or scanner is used and the image is
in color, the raw image is subject to a phenomenon known as “hue
shift” (below, left) wherein the overall color shifts toward a hue, such as red or green. This can occur even after white-balancing the
camera, and the shift differs among brands of cameras, flatbed
scanning devices, and other types of scanners. Post-processing corrects
for color shift so that the final images are better representations
of the original than the raw scans were (below, right).

- Dynamic range. Some important features in the object or sample
may be outside the dynamic range of the image recording device.
These features show up as pure white or pure black in the image
and contain no details (below, left). Often, the researcher will
amplify (or diminish) these features when acquiring the image to
reveal details in other features that are darker or brighter. Without
the ability to alter the relationship of grayscale values (e.g.,
lighten the darker values in relation to the brighter values), the
image is a misrepresentation of the specimen. Again, post-processing
is required to create a truer representation of the original
(below, right).

When and Where Misrepresentation Takes Place
Each example cited previously is an argument for the use of postprocessing
even when instrumentation is used correctly. Still, this
argument is not intended to dismiss the necessity for proper preparation
of specimens or the need for greater education in regard to the
use of imaging equipment and associated instrumentation. These
devices must be used correctly before post-processing.
Also, another factor needs to be considered—how to maintain accurately
represented images for various outputs, including hard copy
(prints and posters), display on computer screens, inclusion in electronic
documents, laptop projection at meetings, and publication.
Originals must be conformed to each output, because each associated
output device or software will modify the appearance of the image to
the degree that the image can become inaccurate.
Thus, the potential for misrepresentation of visual evidence lies in
three areas: when taking an image (acquisition), when correcting that
image (post-processing), and when conforming that image to the output
(conformance).
Much has been written about potential errors for a variety of imaging
systems, including digital cameras, video cameras, scanning beam
systems with photomultiplier (PMT) tubes, and flatbed scanners. In the area of image processing, attention has been drawn either to
instances in which the potential for misrepresenting visual data exists
or to specific instances in which a researcher purposefully altered
visual data to get a desired result. Of course, the potential to alter
visual data is always a possibility, but the system of scientific publication
demands repetition of experiments to corroborate results, making
the system self-correcting. The use of post-processing programs
like Photoshop CS3, when done correctly and when users in all labs
are trained in these methods, results in changes that should be made
versus those that misrepresent visual data. Thus, in this article, references
to changes made to an image are often termed as corrections
versus more widely used terms like alterations or manipulations.
Conforming images to a specific output is an inevitable part of postprocessing.
Changes to the image almost always require adjustments
to grayscale or color tonal values, correction of the contrast, and possibly
a change in both the format and mode of the image.
Note that conformance steps must be taken or images may become
misrepresentations of visual data when reproduced to various outputs
(below, left). While many scientists rely on hardware devices and
software to interpret and interpolate visual data or on color correction
professionals at printing presses, at some point the visual data
will be changed. Ultimately, greater control over reproduction can be
in the hands of scientists (below, right). That kind of control, with the aid of reference material such as the material found in Scientific Imaging with Photoshop, is more likely to prevent perhaps the greatest instance of misrepresentation
in research and what has not been publicly addressed:
reproduction to outputs.
|

Two reproduced images: researcher did not conform color to press output before sending image to the publisher
(left); color was corrected before publishing (right). Science/NSF Visualization Challenge, 1st place, Photography, 2004: image
courtesy of Marna Ericson, PhD.
|
Further Division of Areas into Categories
Images can be divided further into categories, depending on the
intent of the image and the imaging system used for acquisition. Each
group demands its own particular acquisition and post-processing
treatment.
For example, if the visual data is intended for measurement of OD/I,
changes to images acquired from that imaging system must be kept
as close to the original image as possible, with only a few “allowable”
post-processing methods and only the necessary methods for conformance
to outputs (assuming the original was acquired on an imaging
system that results in a linear distribution of grayscale or color
values).
On the other hand, if the image is intended for visualization (in this
sense, a conscious enhancement of visual data to draw attention to
experimental phenomena, such as an image destined for a publication’s
cover or a cartoon model) or any subsequent quantification
(except OD/I), many more changes are allowable. In the case of
quantification, these changes are absolutely necessary for separating
measured features from the surrounding areas, often referred to as
“background.”
The following categories broadly separate the intent and/or the means
for acquiring the image:
- OD/I from flatbed scanners. Images made for measurement of
brightness/darkness levels. These include electrophoretic samples
done in the field of molecular biology (below,left).
- OD/I from camera and scanned beam systems. Images made for
measurement of brightness/darkness levels, or for the measurement
of color, not acquired from flatbed scanners. These include
samples imaged via microscopes, confocal instruments, electron
microscopy, x-ray devices, magnetic resonance imaging, and so on.
- Representation. Images made from any imaging device for purposes
of creating an accurate representation of what was once seen
(bemow, middle).
- Quantification/Visualization. Images made through conscious
alteration of data using pseudocoloring, binarizing, and other
techniques to separate relevant visual information from the background
(below, right).
Author Guidelines
Following specific guidelines for acquisition, post-processing, and
conformance ensures accurate representation of what was once seen
by eye. Because part of the aim of research is to report or publish, the
“gold standard” for ethical guidelines is found in the author guidelines
from major scientific publications.
But the division of visual data into areas and categories will not be
found in author guidelines from major scientific publications, at least
to date. General guidelines are set forth in scientific publications to varying degrees of detail, but often these lack specifics for when and
where potential for misrepresentation may exist. That lack of clarity
leads to varying degrees of interpretation from one investigator
to another and inconsistent responses from journal reviewers and
editors.
Using Standards and References
In addition to when and where potential for misrepresentation exists,
there is also the question of how: How can representative images be
made while preserving a consistent approach to imaging?
The answer isn’t simple. In the best of situations, grayscale and color
values can be objectively determined by fitting the imaging system
(or the image, or the data derived from the image) to an external
standard. Ideally, that standard is a calibrated object with known
values. The imaging system can then be calibrated to the standard,
and presto, all images from that system are also fitted to the known
values. This is typical for situations in which OD/I measurements are
derived from images acquired with self-calibrating, scanner systems.
However, other situations present difficulties. A calibrated standard
may not be available, as in fluorescence imaging. Calibrated standards
may work in an ideal world but not in the real world, such as in colors
that can be defined by standards but do not exist with the same
purity in a sample. The actual specimen may change in grayscale or
color value as a result of preparation techniques and inherent factors,
making a calibrated standard useless. Labeling of specimens may vary
in intensity, making it more important to calibrate to an internal reference
that is part of the specimen versus a calibrated standard. For
these reasons and others, the use of an external, calibration standard
is not always the answer for a consistent approach to imaging, which
is what is desired for reproducibility and correct representations.
A consistent external reference—instead of a standard—may need to
be substituted for a calibrated standard to provide a predictable reference
value against which colors can be corrected or to which energy source intensities or exposure consistencies can be tracked over time.
As in the calibrated standard, the external reference can be included
with the specimen or taken at the beginning or end of an imaging
session.
The reference can be internal: Specimens may have intrinsic values
that can be ratioed against each other, or a consistent grayscale or
color value may be found within the specimen, eliminating the need
for either a calibration standard or external reference.
When no standards or references are available, distributions of grayscale
or color values (histograms) can be matched to either a “perfect,”
reference image, or images can be fit to the dynamic range of the
imaging instrument (when acquiring images) or fit to a common histogram
(post-processing). In that manner, all images are uniform.
As long as a reasoned approach is chosen for the type of specimen and
the intent of the image, representative images can be produced, and
imaging procedures can be duplicated. A summary of the approach to
consistent imaging through the use of calibrated standards, internal
references, or situations in which no standards or references exist is
shown below.

|